Comeco logo

Advantages

  • Operates with standard pH/ORP electrodes
  • High input impedance – over 1013
  • Relay alarm / control output
  • Analog retransmission output isolated from input
  • Pt100, Pt1000 or NTC 1k temperature compensation
  • Fully programmable

Description

PHT20 is a reliable pH/ORP transmitter, enclosed in a small housing for DIN- rail mounting or in box mounting. PHT20 converts the high impedance voltage output of pH or REDOX (ORP) probes into standard current or voltage signal that can be safely sent over longer distances. The transmitter provides the necessary for the normal pH/ORP probe’s operation and has an additional Pt100, Pt1000 or NTC 1k input for automatic temperature compensation. The transmitter provides linear current or voltage output and can also be equipped with a relay alarm/control output. All device parameters (preset conductance ranges, analog output range and direction, relay set point, hysteresis, and direction, temperature compensation parameters, etc.) are user-programmable via the RS232 interface.. Thanks to its wide measurement range, flexibility, programmability, and reliability, the PHT20 transmitter is very suitable for water treatment and processing applications in the food, pharmaceutical, and many other industries.

Additional information

Захранване

10…30 VDC

Релеен изход

MOS gate, None

Корпус

23x90x58 mm, 80x80x60 mm

Монтаж

DIN rail, In box IP65

Интерфейс

RS232

Аналогов изход

0…10 V, 0…20 mA, 2…10 V, 4…20 mA, None, Programmable

Аналогов вход

ORP input, Other, pH input, RTD NTC 1k, RTD Pt100 – 1.385, RTD Pt100 – 1.391, RTD Pt1000 – 1.385

PHT20 transmitter

pH / ORP

Technical Specifications
Input
Input typeStandard pH/ORP electrodes
Measurement rangepH – 0…14, ORP -2000…+2000 mV
Temperature inputPt100, Pt1000, NTC 1k
Temperature compensation0.0…100°C
Accuracy
Total measurement error±0,3% from span (incl. non-linearity)
Temperature drift0.06% from span for 1 °C
Analog output
Output signal0(4)…20 mA, 0(2)…10 V, reversible
Resolution0.025%
Current loadmax. 750 Ω at 20 mA
Voltage loadmax. 2 kΩ at 10 V
Isolation from input1500 VAC for 1 min
Relay output
MOS gate0.1A/60V, optically isolated
Output functionON/OFF, alarms
Set pointconfigurable via interface
Isolation from input1500 VAC for 1 min
Interface
Interface typeRS232 (TTL), requiring special cable (1)
Interface cable type (1)K2, K2U, K12U
Power supply
Supply voltage10…30 VDC
Consumptionmax. 60 mA
Isolation from input1500 VAC for 1 min
Operating conditions
Ambient temperature-10…65 °C
Operating humidity0…85 %RH, non-condensing
Storage temperature-20…85 °C

 

Design and materials‘C’‘Y’
Case materialplasticplastic
Mountingon DIN railfree mounting
Wiringscrew terminalsscrew terminals
Dimensions23x90x58 mm80x80x60 mm
Weightmax.100 gmax.200 g
Protection classIP20IP65

(1) Ordered separately (see ‘Accessories’)